מעקר ואקום פולסים - 352 ליטר, דלת אחת, מחולל קיטור חשמלי מובנה, עגלת העברה חיצונית
Pressure display Analogue pressure gauge for chamber and jacket on the front panel
Temperature display Digital Temperature Display on the front panel (touch screen)
Built-in steam generator 24 kW, working pressure: 0.3MPa
Fittings on built-in steam generator
Safety valve, Analogue pressure gauge, Water level gauge, water level
electric cut out, Automatic air ventilate, Control Switch
Water pump:
Not less than 3 bar should be fitted with a protection against overload and
phase failure.
. Data record Built-in micro printer
certification MDD 93/42 EEC ISO ASMA EMC